Graco Marine Sanitation System 312964C User Manual

312964C  
Tank Level Monitor (TLM)  
EN  
- Monitors tanks levels for oil and anti-freeze mixtures -  
Part No. 256285  
Use with Graco Matrix 3.0 Software only.  
US Patent D484,819  
Important Safety Instructions  
Read all warnings and instructions in this  
manual. Save these instructions.  
Not for use in hazardous locations or in explosive  
atmospheres!  
NOTICE  
Do not over tighten tank level monitor into tank bung!  
Over tightening can cause permanent damage and result  
in inaccurate readings.  
Do not use thread sealant or adhesive! Many of these  
products are chemically incompatible with the PC/ABS  
plastic.  
ti12031a  
Use the Graco-supplied down tube 15U731 only. Do not  
replace down tube with any other tube.  
The Matrix Tank Level Monitor contains an RF device with the following approvals:  
FCC ID: TFB-FREESTAR  
IC: 5969A-FREESTAR  
N14939 - FREESTAR  
ZFMSM-101-1 (CEL) / FS24-100ST (LSR)  
Industry Canada Statement  
The term “IC” before the certification/registra-  
tion number only signifies that the Industry  
Canada technical specifications were met.  
 
Set Up  
Set Up  
There are 8 Network ID's and 8 Transceiver ID's possi-  
ble by changing the position of the dipswitches. The  
eight positions are identified as 1, 2, 3, 4, 5, 6, 7 and 8.  
See FIG. 2 and FIG. 3 on pages 4 and 5.  
Do not install or service this equipment unless you  
are trained and qualified. Installing and servicing  
this equipment requires access to parts which may  
cause fire, explosion, and serious injury if work is  
not performed properly. Read warnings, page 2.  
Setting the S1 and S2 Dipswitches  
(Unless otherwise indicated, for the following instruc-  
tions Refer to FIG. 1 and FIG. 2)  
Do not use the TLM with pressurized tanks.  
1
NOTICE  
4
3
Do not over tighten tank level monitor into tank  
bung! Over tightening can cause permanent dam-  
age and result in inaccurate readings.  
5
Do not use thread sealant or adhesive. Many of  
these products are chemically incompatible with the  
PC/ABS plastic.  
Use the Graco-supplied down tube 15U731 only. Do  
not replace down tube with any other tube.  
7
2
FIG. 1  
Dipswitch Settings  
1. Remove the protective plastic over cover (5) on the  
TLM display that was used for shipping and discard.  
The TLM has two, 4 - position dipswitches labeled S1  
and S2. TLM dipswitches must be set to match those of  
the transceiver the TLM will be communicating with. The  
factory default setting for all TLM’s is (‘1’ ‘1’). The first ‘1’  
refers to the Network ID and the second ‘1’ refers to the  
Transceiver ID.  
2. Remove the four screws (3) holding the tank monitor  
cover (7) in place.  
3. Remove the cover (7).  
Network ID (S1): The RF identification setting  
assigned to a Matrix installation. All compo-  
nents in the system use this same Network ID.  
For example, if one dealership is using Network  
ID (1), the dealership across the street would  
require Network ID (2, 3, 4, 5, 6, 7, or 8) to avoid  
RF interference between the two systems.  
4. Remove the insulating foam (4) to access the circuit  
board.  
5. If installed, remove batteries (1).  
6. Set the S1 and S2 settings to match those of the  
transceiver that this TLM will communicate with  
(FIG. 2 and FIG. 3).  
Transceiver ID (S2): The RF identification set-  
ting assigned to a Matrix Transceiver(s). Matrix  
system components are then assigned to the  
Transceiver’s ID as desired for RF communica-  
tion. For example, if a system required two  
Transceivers, some components would be  
assigned to one Transceiver and other compo-  
nents would be assigned to the second Trans-  
ceiver.  
NOTICE  
Wait at least 30 seconds after the dipswitch settings  
are made before installing the batteries. If you do not  
wait the 30 seconds, the software will not recognize  
the new settings.  
312964C  
3
 
   
Set Up  
7. Install two, 9-volt alkaline batteries (1). Be sure that  
the batteries fully engage the mounting clips by  
pushing on the bottom of each battery with your  
thumb.  
9. Replace cover (7) and secure it with the four screws  
(3).  
Make sure the cover screws are tightened securely  
(18-22 in-lb) to avoid water leakage into the TLM  
electronics. If a torque wrench is not used, verify  
there are no gaps under the screw heads and no  
gaps under the cover flange. This will ensure  
proper compression of the gasket for a water-tight  
seal.  
It will take about 30 seconds for the monitor to dis-  
play information after pressing the Display button  
on the initial power-up. This time will decrease  
thereafter to a few seconds  
8. Replace insulating foam (4).  
Make sure the gasket (2) is not damaged and is in  
the correct location.  
S1 sets NETWORK ID  
S2 sets TRANSCEIVER ID  
All dipswitches in down  
or off position.  
on  
S2  
4
1
2
3
S1  
All dipswitches  
to right or off  
position.  
ti12007a  
FIG. 2  
4
312964C  
 
 
Set Up  
Dipswitch Setting for Network ID and Transceiver ID  
See FIG. 3 for dipswitch setting configurations.  
b
The batteries must be disconnected and removed before changing dipswitch settings or the TLM will  
not be able to communicate with the PC software.  
S2 Transceiver ID  
3
4
1 - default  
2
on  
on  
on  
on  
S2  
S2  
S2  
S2  
S2  
S2  
S2  
4
4
4
1
2
2
1
2
2
4
4
4
3
3
3
3
1
2
1
2
2
3
3
7
6
5
8
on  
on  
on  
on  
S2  
1
1
4
3
4
1
2
1
3
S1 Network ID  
8
3
7
2
4
5
6
1 - default  
S1  
S1  
S1  
S1  
S1  
S1  
S1  
S1  
FIG. 3  
312964C  
5
 
 
Set Up  
Registering the TLM  
Graco recommends registering TLM prior to installa-  
tion.  
See FIG. 4 for display messages you will see during  
Registration Mode.  
The TLM operating parameters are controlled by the  
Matrix PC Software and setup by the System Adminis-  
trator. See the Matrix 3 Software instruction manual for  
the PC setup instructions.  
To register the TLM:  
1. Set TLM to correct NETWORK ID and TRANS-  
CEIVER ID (see Dipswitch Settings, page 5).  
A
2. Press and hold the Display button (A, FIG. 4) until  
display (B) says “Register Mode”.  
B
3. Display reads “Registration Complete” when the  
TLM is registered with the PC software.  
ti12032a  
FIG. 4  
6
312964C  
 
TLM Display  
TLM Display  
Fluid Volume -Volume of fluid remaining in the tank,  
based on the tank geometry defined during setup. The  
TLM can be programmed to display in either gallons or  
liters during programming of the TLM.  
TLM ID - Unique number assigned to the TLM to distin-  
guish it from other TLM’s in the system.  
Battery Life - Remaining life of the unit’s batteries. Sys-  
tem Administrator on Matrix PC Software determines  
when Low Battery Warning message is generated.  
Vertical tanks use this capacity figure to calculate  
tank volume. Obround and cylindrical tanks use the  
tank dimensions to calculate tank volume.  
Battery life is dependent on the number of readings  
taken per day.  
No RF Signal - Displays when the TLM is not receiving  
an RF ACK signal from the PC transceiver. If the tank  
level monitor is receiving an ACK signal or the signal is  
re-established, this screen will not appear.  
Network ID and Transceiver ID - Shows the Network  
ID and Transceiver ID TLM is using (also see page 3).  
If the message “No RF Signal” appears, the fluid volume  
and battery life data is not transmitted or updated at the  
PC. This message may take up to 15 seconds to display.  
Firmware - Current version of operating software con-  
trolling TLM operation.  
Fluid Level - Depth of fluid remaining in the tank pro-  
vided in either inches or centimeters as defined during  
setup.  
No PC Signal - Displays when the TLM is not receiving  
a response from the Matrix PC software (i.e, PC is off or  
software is not installed). If the TLM receives a response  
from the PC, this information will not appear.  
If the message “NO PC SIGNALappears, the fluid vol-  
ume and battery life data will not be updated at the PC.  
This message may take up to 15 seconds to display.  
312964C  
7
 
TLM Registration Process  
TLM Registration Process  
Press & Hold  
DISPLAY button  
TLM ID  
Battery  
NW ID  
TX ID  
REGISTER  
MODE  
No Ack received from  
Transceiver  
Ack received from  
Transceiver  
TLM ID  
TLM ID  
Display Reads  
No RF Signal  
VERSION  
xx.xx.xx  
VERSION  
xx.xx.xx  
BLANK  
BLANK  
PC Responds with  
Registration Ack  
Message  
PC doesn’t Respond  
REGISTER  
COMPLETE  
Display Reads  
No PC Signal  
FIG. 5  
8
312964C  
 
 
Installation  
Installation  
NOTICE  
Do not over tighten tank level monitor into tank  
bung! Over tightening can cause permanent dam-  
age and result in inaccurate readings.  
Do not use thread sealant or adhesive! Many of  
these products are chemically incompatible with the  
PC/ABS plastic.  
Use the Graco-supplied down tube 15U731 only. Do  
not replace down tube with any other tube.  
TLM must be mounted with-  
in 2° from perpendicular to  
the surface of the fluid.  
2°  
2°  
1. Remove the bung fitting and screw in the TLM  
hand-tight. DO NOT use a wrench to tighten as this  
may damage the TLM. Do not use thread lock adhe-  
sive as this may damage the TLM.  
FIG. 6  
2. The TLM must be mounted within 2 degrees from  
perpendicular to the surface of the fluid. A level that  
measures degrees should be used. Place the level  
on the top of the pipe fitting on the tank. If outside  
the 2 degree specification, the fitting should be  
changed.  
NOTICE  
Do not install the TLM in the tank’s fill port. Repeated  
removal of the TLM will damage the unit and void the  
warranty.  
Graco recommends that the height of the tank fitting  
should not exceed 1/2 in. (1.27 cm). The fitting  
height can be up to 2 in. (5.08 cm), provided the fit-  
ting is within 2 degrees of the surface of the fluid.  
Also, the taller the tank, the more critical the 2  
degree specification becomes. See FIG. 6.  
Port (D) is not to be used as a fill hole for oil deliveries.  
See FIG. 7.  
The tank’s breather port might be the only available  
location for mounting the TLM. The horizontal 2 in. npt  
port (D) can be used to accommodate the breather. The  
breather should be mounted in an upright position using  
a 90° elbow fitting.  
NOTICE  
The TLM will not operate correctly if tilted more than  
2° from the surface of the tank liquid.  
Do not use the horizontal port for routing pressure  
relief return tubes. This can result in inaccurate  
readings.  
The TLM will not read properly if:  
-
Fluid in the tank is agitated (i.e. filling a tank  
with oil or anti-freeze). Be sure to take all TLM  
readings when the fluid in the tank is calm.  
-
-
TLM is mounted next to the tank wall or in a cor-  
ner.  
D
If there is any structural bracing or obstruction  
between the TLM and the surface of fluid in  
tank.  
-
If the down tube is not the Graco-supplied down  
tube.  
ti12034a  
FIG. 7  
312964C  
9
 
 
Viewing Data  
Viewing Data  
After a tank is refilled from an oil distributor or a  
waste oil tank is emptied by a waste oil service  
provider, press the Display button on the TLM to  
ensure the TLM will maintain the most current  
tank volume status. If button is not pushed, the  
TLM will automatically read correctly at the next  
scheduled tank reading.  
Daylight Saving Time (DST) Change  
B
The PC software automatically changes the PC time  
when daylight saving time changes occur (Spring and  
Fall).  
A
FIG. 8  
The first TLM reading will be 1 hour earlier or later  
(depending on the Spring or Fall daylight saving  
time change), but all subsequent TLM reading will  
be accurate.  
Press and release the Display button (A). Data appears  
on display (B); each screen lasting a few seconds before  
going on to the next. See FIG. 8.  
Tank Level data can be viewed remotely at the PC,  
using the Matrix software.  
TLM ID  
03000001  
First Display  
Battery  
Second Display  
100%  
Nw ID - A  
Third Display  
Tx ID - A  
36‚  
Fluid  
Fourth Display  
Fifth Display  
(can program for liters)  
Gallons  
100  
Screen display when  
No RF  
no RF ACK signal is  
received after 10-12  
seconds  
Signal  
Displays when PC  
doesn’t respond after  
10-12 seconds  
NO PC  
Signal  
FIG. 9:  
10  
312964C  
 
 
Troubleshooting  
Troubleshooting  
Problem  
Cause  
Solution  
Brand new monitor’s dis- Unit is shipped with protective coat- Remove protective coating.  
play is very dim. ing over display.  
Monitor displays “Invalid The TLM reading is outside of the  
Verify the programmed parameters and re-program the TLM if  
necessary.  
Reading”.  
programmed size parameters.  
Monitor will not register.  
Transceiver is not powered-up.  
Verify transceiver is powered-up  
Attempt to program while red lights Wait until red lights on transceiver go blank before attempting  
on transceiver are lit. to register.  
Transceiver NET ID and TRANS ID Ensure NET ID and TRANS ID are set for appropriate commu-  
not set correctly  
nication cable.  
TLM dipswitches settings do not  
match transceiver settings.  
Verify settings with Transceiver (see page. 5 for dipswitch set-  
ting instructions).  
Weak or dead batteries.  
Replace batteries. See Replacing Batteries, page 13.  
Microprocessor not completely shut After you have removed old batteries, wait 30 seconds to  
down before installing new batteries. ensure the microprocessor has completely shut down, before  
installing new batteries.  
USB cable or RS-422 not connected Be sure cable are properly connected between Transceiver  
and PC  
Monitor has intermittent  
RF communication  
Out of RF range.  
RF obstruction.  
Reposition Transceiver until TLM has good RF communication  
or add a transceiver.  
Weak or dead batteries.  
Out of RF range.  
RF obstruction.  
Replace batteries. See Replacing Batteries, page 13.  
Monitor is not reporting  
scheduled readings.  
Reposition Transceiver until TLM has good RF communication.  
Two TLMs are programmed to the  
same address.  
Verify that each TLM is correctly programmed to a unique  
address.  
Weak or dead batteries.  
Replace batteries. See Setting the S1 and S2 Dipswitches,  
Microprocessor not completely shut After you have removed old batteries, wait 30 seconds to  
down before installing new batteries. ensure the microprocessor has completely shut down, before  
installing new batteries.  
Tank Level Monitor was not associ- Reregister Tank Level Monitor and associate Tank Level Moni-  
ated with profile after scheduled  
times were entered into software.  
tor with profile ID.  
Reading is scheduled when pump is Ensure readings are scheduled at times that the pump is not  
operating.  
operating.  
Clock was changed on PC but the  
The monitor’s internal clock is synchronized with the PC clock  
tank level monitor was not resychro- upon register or reading. Manipulating the scheduled reporting  
nized.  
time by changing the PC clock will cause a false indication that  
the monitor is not reporting at the scheduled reporting times.  
Monitor will not take read- Weak or dead batteries.  
ing when Display button  
is pressed.  
Replace batteries. See Replacing Batteries, page 13.  
Batteries are not correctly seated.  
Ensure the batteries fully engage the mounting clips by push-  
ing on the bottom of each battery with your thumbs.  
Microprocessor not completely shut After you have removed old batteries, wait 30 seconds to  
down before installing new batteries. ensure the microprocessor has completely shut down, before  
installing new batteries.  
Monitor display is cracked.  
PC is not running  
Replace monitor display.  
Turn on PC. In order for TLM readings to be logged, PC must  
be running.  
312964C  
11  
 
Troubleshooting  
Problem  
Cause  
Solution  
Monitor readings are  
inaccurate.  
Tank geometry incorrectly defined.  
See Tank Level Monitor Software Guide or PC Software Guide  
for details.  
Tank Level Monitor has not been  
updated with latest adjustments  
made within The profile ID on the  
PC.  
Manually push DISPLAY button on TLM  
Two TLMs are programmed to the  
same address.  
Verify that each TLM is correctly programmed to a unique  
address.  
Fluid surface is moving while reading Ensure pump is not operating and that nothing is disturbing the  
is being taken.  
surface of the fluid during readings.  
Pipe adapters installed in the tank  
bung.  
Ensure there are no adapters installed in the tank bung.  
Tank Level Monitor is not perpendic- Level tank so that it is perpendicular to top of the fluid and/or  
ular to top of fluid.  
realign Tank Level Monitor.  
Inside tank obstruction.  
Install in different tank bung to avoid tank obstruction.  
12  
312964C  
 
Service  
Service  
Once the Tank Level Monitor has been installed, no additional maintenance or service is necessary, with the excep-  
tion of replacing batteries.  
Replacing Batteries  
Always replace both batteries with two new 9-volt alkaline batteries whenever you change batteries in the TLM.  
Sparking can occur when changing batteries. Only  
replace the battery in a non-hazardous location,  
away from flammable fluids or fumes.  
1. Remove the four cover screws (3) and cover (7).  
2. Disconnect and remove both batteries (1) from TLM  
and discard batteries in approved battery disposal  
container.  
3. Replace discarded batteries with two new batteries.  
Be sure the batteries fully engage the mounting  
clips by pushing on the bottom of each battery with  
your thumb.  
4. Replace the cover (7) and screws (3). Tighten the  
four cover screws.  
Make sure the cover screws are tightened  
securely (18-22 in-lb) to avoid water leakage  
into the TLM electronics. If a torque wrench is  
not used, verify there are no gaps under the  
screw heads and no gaps under the cover  
flange. This will ensure proper compression of  
the gasket for a water-tight seal.  
After changing batteries it is not necessary to  
reprogram the TLM. However, the Display but-  
ton should be pressed to synchronize the  
TLM’s internal clock with the PC clock. If not  
done, the TLM will not report at the correct  
scheduled times.  
312964C  
13  
 
 
Parts  
Parts  
1
2
4
3
8
1
6
7
2
5
9
10  
ti12033a  
1
Remove protective cover used for shipping before  
programming.  
2
Do not remove. This is required to maintain intrinsic safety approval.  
Ref.  
No.  
Part No. Description  
Qty  
1
2
3
4
5
6
7
8
9
BATTERY, 9 volt (purchased locally)  
126752 GASKET, sensor, ultrasonic  
117467 SCREW, self tapping, HI-LO, #10-16 x 9/16  
117743 FOAM, insulator  
117259 ENCLOSURE  
15T814 LABEL, overlay  
15A962 COVER  
117469 HOLDER, battery  
15U731 DOWN TUBE  
2
1
4
1
1
1
1
2
1
10 119391 PLUG  
14  
312964C  
 
Technical Data  
Technical Data  
Ultrasonic Tank Depth Measurement range  
Fluid Level Measurement Accuracy  
Mounting  
0 - 30 ft. (0 - 9 m) Not for use in pressurized tanks.  
+/- 0.5%  
Standard 2 in. (npt) bung.  
Height above tank for TLM and tube mount- 15 in. (30.48 cm).  
ing  
TLM Mounting Tube and Tank Vent  
Graco recommends venting the tank separately from the TLM. If not  
possible, the TLM mounting and vent function can be combined if  
required for tanks with one hole in tank. Be sure TLM is not mounted  
close to tank edges or corner.  
Weight (with batteries installed)  
RF Communication  
2.34lb. (1.06 g)  
2.4 GHz  
Operating Temperature Range  
-22° F to 185° F (- 30° C to 85° C) Note: Display will not function  
below 32°F (0°C).  
Storage Temperature Range  
Batteries  
-40°F to 185°F (-40°C to 85°C)  
Two 9V alkaline  
1.5 years  
Battery Life  
Enclosure  
IP65  
RF Communication  
2.4 GHz  
Obstructed RF Communication Range  
(based on building construction and RF envi-  
ronment)  
250-300 ft (76.2 - 91.0 m)  
Tank Geometry  
Vertical walled tanks, cylindrical tanks, and obround tanks.  
•Vertical Tanks  
Maximum Volume 999,999 gallons or liters  
Maximum Height 30 ft. (360 in.)  
•Cylindrical Tanks  
Maximum Volume 999,999 gallons or liters  
Maximum Diameter 30 ft. (360 in.)  
Maximum Length 30 ft. (360 in.)  
•Obround Tanks  
Maximum Volume 999,999 gallons or liters  
Maximum Height 30 ft. (360 in.)  
Maximum Width 30 ft. (360 in.)  
Maximum Length 30 ft. (360 in.)  
Approvals  
FCC, Industry Canada (IC), C-Tick, CE  
NOTE: FCC, C-Tick and IC approvals are for the RF device contained in the Matrix Tank Level Monitor (TLM).  
312964C  
15  
 
Dimensions  
Dimensions  
A
B
9.1 in (231 mm)  
4.9 in. (124 mm)  
B
A
1-1/2 in. npt  
2 in. npt  
ti12031a  
16  
312964C  
 
Tank Geometry  
Tank Geometry  
Vertical Tank:  
Maximum Volume = 999, 999 gallons or liters.  
Tank walls must be uniformly vertical from empty level to full  
Maximum Height  
30 ft. (360 in.)  
ti12693a  
FIG. 10  
312964C  
17  
 
Tank Geometry  
Cylindrical Tank:  
Maximum Volume = 999, 999 gallons or liters.  
Tank end walls must be flat; they cannot be any other shape, including belled.  
Maximum Height  
30 ft. (360 in)  
Maximum Length  
30 ft. (360 in)  
ti12695a  
FIG. 11  
18  
312964C  
 
Tank Geometry  
Obround Tank:  
Maximum Volume = 999, 999 gallons or liters.  
Tank end walls must be flat; they cannot be any other shape including belled.  
Maximum Height  
30 ft. (360 in)  
Maximum Width  
30 ft. (360 in.)  
Maximum Length  
30 ft. (360 in)  
ti12694a  
FIG. 12  
312964C  
19  
 
Graco Extended Tank Level Monitor Warranty  
Graco warrants all equipment manufactured by Graco and bearing its name to be free from defects in material and workmanship  
on the date of sale to the original purchaser for use. With the exception of any special, extended, or limited warranty published by  
Graco, Graco will, for a period of twenty four months from the date of sale, repair or replace any part of the equipment determined  
by Graco to be defective. This warranty applies only when the equipment is installed, operated and maintained in accordance with  
Graco's written recommendations.  
This warranty does not cover, and Graco shall not be liable for general wear and tear, or any malfunction, damage or wear caused  
by faulty installation, misapplication, abrasion, corrosion, inadequate or improper maintenance, negligence, accident, tampering,  
or substitution of non-Graco component parts. Nor shall Graco be liable for malfunction, damage or wear caused by the  
incompatibility of Graco equipment with structures, accessories, equipment or materials not supplied by Graco, or the improper  
design, manufacture, installation, operation or maintenance of structures, accessories, equipment or materials not supplied by  
Graco.  
This warranty is conditioned upon the prepaid return of the equipment claimed to be defective to an authorized Graco distributor  
for verification of the claimed defect. If the claimed defect is verified, Graco will repair or replace free of charge any defective parts.  
The equipment will be returned to the original purchaser transportation prepaid. If inspection of the equipment does not disclose  
any defect in material or workmanship, repairs will be made at a reasonable charge, which charges may include the costs of parts,  
labor, and transportation.  
THIS WARRANTY IS EXCLUSIVE, AND IS IN LIEU OF ANY OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING BUT  
NOT LIMITED TO WARRANTY OF MERCHANTABILITY OR WARRANTY OF FITNESS FOR A PARTICULAR PURPOSE.  
Graco's sole obligation and buyer's sole remedy for any breach of warranty shall be as set forth above. The buyer agrees that no  
other remedy (including, but not limited to, incidental or consequential damages for lost profits, lost sales, injury to person or  
property, or any other incidental or consequential loss) shall be available. Any action for breach of warranty must be brought within  
two (2) years of the date of sale Graco makes no warranty, and disclaims all implied warranties of merchantability and fitness for a  
particular purpose in connection with accessories, equipment, materials or components sold but not manufactured by Graco.  
These items sold, but not manufactured by Graco (such as electric motors, switches, hose, etc.), are subject to the warranty, if any,  
of their manufacturer. Graco will provide purchaser with reasonable assistance in making any claim for breach of these warranties.  
In no event will Graco be liable for indirect, incidental, special or consequential damages resulting from Graco supplying  
equipment hereunder, or the furnishing, performance, or use of any products or other goods sold hereto, whether due to a breach  
of contract, breach of warranty, the negligence of Graco, or otherwise.  
FOR GRACO CANADA CUSTOMERS  
The parties acknowledge that they have required that the present document, as well as all documents, notices and legal  
proceedings entered into, given or instituted pursuant hereto or relating directly or indirectly hereto, be drawn up in English. Les  
parties reconnaissent avoir convenu que la rédaction du présente document sera en Anglais, ainsi que tous documents, avis et  
procédures judiciaires exécutés, donnés ou intentés à la suite de ou en rapport, directement ou indirectement, avec les  
procedures concernées.  
TO PLACE AN ORDER, contact your Graco distributor or call to identify the nearest distributor.  
Phone: 612-623-6928 or Toll Free: 1-800-533-9655, Fax: 612-378-3590  
All written and visual data contained in this document reflects the latest product information available at the time of publication.  
Graco reserves the right to make changes at any time without notice.  
Original Instructions. This manual contains English. MM 312964  
Graco Headquarters: Minneapolis  
International Offices: Belgium, China, Japan, Korea  
GRACO INC. P.O. BOX 1441 MINNEAPOLIS, MN 55440-1441  
Copyright 2008, Graco Inc. is registered to I.S. EN ISO 9001  
Revised September 2012  
 

Grandstream Networks Dishwasher GXV300X User Manual
Hand Held Products Scanner IT4410 User Manual
Harman Kardon Stereo Receiver 374 User Manual
Hawking Technology Switch CS152F User Manual
Hearth and Home Technologies Stove 823 0170 MBK User Manual
Hotpoint Convection Oven SH 108 CX S User Manual
Hotpoint Washer AQ113DA 697 S User Manual
HP Hewlett Packard Network Card 5300A User Manual
HP Hewlett Packard Printer 1005 Series User Manual
HP Hewlett Packard Weather Radio 16500A User Manual